OmniPod Needle Mechanism May Fail to Deploy

Arden's Day Blog

Arden's Day is a type I diabetes care giver blog written by author Scott Benner. Scott has been a stay-at-home dad since 2000, he is the author of the award winning parenting memoir, 'Life Is Short, Laundry Is Eternal'. Arden's Day is an honest and transparent look at life with diabetes - since 2007.

type I diabetes, parent of type I child, diabetes Blog, OmniPod, DexCom, insulin pump, CGM, continuous glucose monitor, Arden, Arden's Day, Scott Benner, JDRF, diabetes, juvenile diabetes, daddy blog, blog, stay at home parent, DOC, twitter, Facebook, @ardensday, 504 plan, Life Is Short, Laundry Is Eternal, Dexcom SHARE, 生命是短暂的,洗衣是永恒的, Shēngmìng shì duǎnzàn de, xǐyī shì yǒnghéng de

OmniPod Needle Mechanism May Fail to Deploy

Scott Benner

The post title has been edited to remove the phrase "voluntary recall" because I mistakenly used the phrase. This is not a recall, voluntary or otherwise - it is a Field Safety Notification. My apologies to all. - Best, Scott

from Insulet


Insulet Corporation
November 2, 2015
URGENT: Field Safety Notification
OmniPod® Insulin Management System
Certain OmniPod Lots Specified Below
 
Dear Valued Insulet Customer,
Insulet Corporation, the manufacturer of the OmniPod Insulin Delivery System, is committed to keeping you and your healthcare professionals up-to-date in the event there are any issues that arise related to our products.

As part of our product quality monitoring process, we have identified that 15 lots of OmniPod which were distributed in the U.S. had a slight increase in the reported cases in which the Pod’s needle mechanism failed to deploy or there was a delay in the deployment of the needle mechanism. The reported incidence of this product issue in the affected lots is approximately 1%-2%.  Once we recognized this, we corrected the manufacturing process and implemented additional inspection steps.

No serious injuries or deaths have been reported in patients using OmniPod devices from the affected lots.

How do I know if I have affected product? 
This Field Safety Notification affects only the Pods and does not affect the OmniPod Personal Diabetes Manager (PDM). The slight increase was identified in the following lots of Pods:


The lot number is located on the Pod tray lid label, the side of the Pod and on the end of the box of Pods.  

To open a PDF version of the Insulet Announcement - click

To open a PDF version of the Insulet Announcement - click


What is the risk? 
In the event a needle mechanism fails to deploy, the needle will not be inserted and insulin delivery will not begin. The interruption of insulin delivery may cause elevated blood glucose (hyperglycemia), which, if left untreated, can result in diabetic ketoacidosis (DKA). If you believe you have successfully activated your Pod and you experience unexpected elevated blood glucose levels, please consult your healthcare professional.

Upon activation, how do I know if the needle mechanism deployed? 
After you press the Start button on the PDM, you should hear a few soft clicks and then a louder click from the Pod indicating that the needle mechanism has deployed. If you do not hear this within a few seconds after pressing Start, the Pod has failed to deploy. 
You should also feel the insertion of the needle mechanism deployment. The PDM will indicate that the Pod is active and prompt you to check to ensure the cannula is properly inserted (Figure 5-23 in my PDF).  For more information, please see the User Guide.
You should always check the infusion site after insertion to ensure that the cannula was properly inserted. 

When you see the pink slide insert in this position, it means that the cannula is inserted (See figure in my PDF).

The PDM will automatically remind you to check your blood glucose 1.5 hours after each Pod change.  If the cannula is not properly inserted, hyperglycemia may result.
If you experience unexpected elevated blood glucose levels, change your Pod and contact your healthcare provider.You may also call Customer Care at 1-855-407-3729 if you have any questions regarding this Field Safety Notification. 

This voluntary action is being taken by Insulet Corporation with the knowledge of the U.S. Food and Drug Administration (FDA). Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA’s MedWatch Adverse Event Reporting program either online or by regular mail or by fax.
 
We appreciate how you depend on us and sincerely regret any inconvenience this may cause you. We are focused on delivering the highest level of product quality and your complete satisfaction is our top priority.


Sincerely,
Michael Spears
Vice President, Quality, Regulatory & Clinical Affairs
Insulet Corporation
 

Announcement can be found on the Insulet website with this link.