RECALL: Medtronic Infusion Sets

Arden's Day Blog

Arden's Day is a type I diabetes care giver blog written by author Scott Benner. Scott has been a stay-at-home dad since 2000, he is the author of the award winning parenting memoir, 'Life Is Short, Laundry Is Eternal'. Arden's Day is an honest and transparent look at life with diabetes - since 2007.

type I diabetes, parent of type I child, diabetes Blog, OmniPod, DexCom, insulin pump, CGM, continuous glucose monitor, Arden, Arden's Day, Scott Benner, JDRF, diabetes, juvenile diabetes, daddy blog, blog, stay at home parent, DOC, twitter, Facebook, @ardensday, 504 plan, Life Is Short, Laundry Is Eternal, Dexcom SHARE, 生命是短暂的,洗衣是永恒的, Shēngmìng shì duǎnzàn de, xǐyī shì yǒnghéng de

RECALL: Medtronic Infusion Sets

Scott Benner

From Medtronic

Explanation of the Issue
Medtronic has become aware of recent reports of potential over-delivery of insulin shortly after an infusion set change. Over-delivery of insulin can cause hypoglycemia and in extreme cases, death. Medtronic has received reports of hypoglycemia requiring medical intervention potentially related to this issue.

Our investigation has shown this can be caused by fluid blocking the infusion set membrane during the priming/fill-tubing process. A membrane blocked by fluid most likely occurs if insulin, alcohol, or water is spilled on top of the insulin reservoir which then could prevent the infusion set from working properly. Infusion sets currently being shipped by Medtronic contain a new and enhanced membrane material that significantly reduces this risk.

To see if your infusion sets are part of the recall please go to this page supplied by Medtronic

https://www.medtronicdiabetes.com/customer-support/product-and-service-updates/notice7-letter

*****
Further Press release